MHRA alerts & safety notifications
Search medicines recalls, device safety information, field safety notices, patient safety alerts, drug safety updates, and safety round-ups from the UK Medicines and Healthcare products Regulatory Agency (MHRA).
GOV.UK MHRA directory refreshed hourly — enable email digests in your PharmAnalyser notifications settings.
Official MHRA drug and device safety notifications, indexed for fast pharmacy and GP practice research.
Filter by alert type, recall class, pharmacy relevance, therapeutic area, and free-text search.
Open any row for full alert text, attachments, and change history — without loading the entire directory.
Figures reflect the ingested GOV.UK directory; always follow the published alert for operational decisions.
Source: GOV.UK drug and device alerts. PharmAnalyser indexes the directory for research — not a live regulatory system of record.
Matching filters
2,317
Current search and filter selection
Class 1 in view
28
Highest urgency recalls in current results
Last 7 days
0
Recently updated in current results
Pharmacy-tagged
559
Across the full indexed directory
Alert type
Recall class
Class is set only when MHRA publishes it in the title — unclassified alerts are not treated as Class 1.
2,317 alerts match your filters
Alert
Summary
Updated
Actions
Medicines RecallClass 3
Class 3 Medicines Recall: Norgine Limited, MOVICOL Plain 13.7g powder for oral solution, EL(26)A/41
PharmacyGP
Norgine Limited is recalling specific batches of product as a precautionary measure due to some units containing the incorrect amounts of active ingredients.
Class 4 Medicines Defect Notification: Neuraxpharm UK Limited, Levetiracetam Neuraxpharm granules for oral solution in sachet, EL(26)A/40
PharmacyGP
Neuraxpharm UK Limited have informed the MHRA that a review of the approved Patient Information Leaflet (PIL) has identified discrepancies between the current Neuraxpharm product safety information and the corresponding reference product information.
Class 4 Medicines Defect Notification: B. Braun Melsungen AG, Meropenem 1000 mg powder and solvent for solution for infusion, EL(26)A/39
PharmacyGP
B. Braun Melsungen AG has notified the MHRA that errors have been identified in the Patient Information Leaflet (PIL) included in some batches of Meropenem 1000 mg powder and solvent for solution for infusion.
Radiofrequency (RF) Safety Considerations for Wireless Cardiotocography (CTG) Monitoring Systems (DSI/2026/012)
The MHRA has conducted a review into the use of wireless cardiotocography (CTG) systems that use radiofrequency (RF) telemetry channels. Signal crossover, erroneous CTG traces and signal dropouts have been identified as patient safety risks associated with these devices.
Class 4 Medicines Defect Notification: Martindale Pharmaceuticals Ltd, Clobazam Martindale Pharma 5 mg/5 mL and 10 mg/5 mL Oral Suspension 150ml, EL(26)A/38
PharmacyGP
Martindale Pharmaceuticals has informed the MHRA of an error with the linear barcode of some batches of Clobazam 5mg/5mL and 10mg/5mL Oral Suspension 150mL.
National Patient Safety Alert: Patient hoists and slings (all types): risk of death and serious harm from falls (NatPSA/2026/005/MHRA)
Fatal and serious harm continues to occur when patients fall from hoists or slings during transfers and repositioning. A review of recent incidents with a fatal outcome and wider surveillance data shows an average of two deaths per year since 2015.
Filters should be used during the administration of Parenteral Nutrition for patients in all care settings
Administration of parenteral nutrition without a filter has been associated with a number of adverse incidents including an embolism which resulted in a fatal outcome.
Filters should be used during the administration of Parenteral Nutrition for patients in all care settings DSI/2026/011
GP
Administration of parenteral nutrition without a filter has been associated with a number of adverse incidents including an embolism which resulted in a fatal outcome.
Cobalt-chrome modular neck hip replacements: risk of metal-wear effects and revision surgery (DSI/2026/010)
GP
An MHRA investigation has found increased risks of wear and corrosion associated with multiple cobalt-chrome (Co-Cr) modular neck hip stem components, including a potential increased occurrence of revision surgery due to adverse local tissue reaction (ALTR).
NGPod pH testing device: quarantine and dispose of all devices (DSI/2026/009)
The NGPod handheld device and NGPod sensor, used to test pH to aid the placement of nasogastric tubes, is being withdrawn from the UK market. The manufacturer has ceased trading in the UK. Users should quarantine all remaining stock and dispose of all NGPod devices.
General SurgeryGi Hepatology Pancreatic Disorders1Full detail
Rectal catheters sold for personal use (gas and colic relievers): Do not use in infants, including for the treatment of colic, gas-related discomfort or constipation (DSI/2026/008)
GP
Do not use rectal catheters sold for personal use (gas and colic relievers) in infants, including for colic, gas-related discomfort or constipation. Stop using these products immediately and dispose of any unused devices.
General practiceGi Hepatology Pancreatic Disorders1Full detail
Class 2 Medicines Recall: Zentiva Pharma UK Limited, Fingolimod Zentiva 0.5 mg Capsules, EL(26)A/37
PharmacyGP
Zentiva UK Pharma Limited are recalling one batch of Fingolimod Zentiva 0.5 mg Capsules (BN 4L01372H) as a precautionary measure due to a potential risk of contamination with metal particles.
National Patient Safety Alert: ResMed Astral 100 and 150 Ventilators: Potential for Patient Harm due to Unexpected Interruption of Ventilation Therapy (NatPSA/2026/004/MHRA)
ResMed has issued a Field Safety Notice regarding Astral 100 and 150 ventilators and associated Printed Circuit Board Assembly spare parts manufactured prior to October 2024. A fault in the device can cause unexpected interruption of ventilation therapy.
Devices supplied without valid UKCA/CE conformity markings or certification: remove from use and place in quarantine (DSI/2026/007)
PharmacyGP
The MHRA has become aware of a number of medical devices being supplied on the UK market that do not comply with the requirements of the UK Medical Device Regulations 2002 (UK MDR 2002) or with the EU Medical Devices Regulation (EU) 2017/745 (EU MDR).
Class 3 Medicines Recall: Norgine Limited, MOVICOL Plain 13.7g powder for oral solution, EL(26)A/41
PharmacyGP
Norgine Limited is recalling specific batches of product as a precautionary measure due to some units containing the incorrect amounts of active ingredients.
Class 4 Medicines Defect Notification: Neuraxpharm UK Limited, Levetiracetam Neuraxpharm granules for oral solution in sachet, EL(26)A/40
PharmacyGP
Neuraxpharm UK Limited have informed the MHRA that a review of the approved Patient Information Leaflet (PIL) has identified discrepancies between the current Neuraxpharm product safety information and the corresponding reference product information.
Class 4 Medicines Defect Notification: B. Braun Melsungen AG, Meropenem 1000 mg powder and solvent for solution for infusion, EL(26)A/39
PharmacyGP
B. Braun Melsungen AG has notified the MHRA that errors have been identified in the Patient Information Leaflet (PIL) included in some batches of Meropenem 1000 mg powder and solvent for solution for infusion.
Radiofrequency (RF) Safety Considerations for Wireless Cardiotocography (CTG) Monitoring Systems (DSI/2026/012)
The MHRA has conducted a review into the use of wireless cardiotocography (CTG) systems that use radiofrequency (RF) telemetry channels. Signal crossover, erroneous CTG traces and signal dropouts have been identified as patient safety risks associated with these devices.
Class 4 Medicines Defect Notification: Martindale Pharmaceuticals Ltd, Clobazam Martindale Pharma 5 mg/5 mL and 10 mg/5 mL Oral Suspension 150ml, EL(26)A/38
PharmacyGP
Martindale Pharmaceuticals has informed the MHRA of an error with the linear barcode of some batches of Clobazam 5mg/5mL and 10mg/5mL Oral Suspension 150mL.
National Patient Safety Alert: Patient hoists and slings (all types): risk of death and serious harm from falls (NatPSA/2026/005/MHRA)
Fatal and serious harm continues to occur when patients fall from hoists or slings during transfers and repositioning. A review of recent incidents with a fatal outcome and wider surveillance data shows an average of two deaths per year since 2015.
Filters should be used during the administration of Parenteral Nutrition for patients in all care settings
Administration of parenteral nutrition without a filter has been associated with a number of adverse incidents including an embolism which resulted in a fatal outcome.
Filters should be used during the administration of Parenteral Nutrition for patients in all care settings DSI/2026/011
GP
Administration of parenteral nutrition without a filter has been associated with a number of adverse incidents including an embolism which resulted in a fatal outcome.
Cobalt-chrome modular neck hip replacements: risk of metal-wear effects and revision surgery (DSI/2026/010)
GP
An MHRA investigation has found increased risks of wear and corrosion associated with multiple cobalt-chrome (Co-Cr) modular neck hip stem components, including a potential increased occurrence of revision surgery due to adverse local tissue reaction (ALTR).
NGPod pH testing device: quarantine and dispose of all devices (DSI/2026/009)
The NGPod handheld device and NGPod sensor, used to test pH to aid the placement of nasogastric tubes, is being withdrawn from the UK market. The manufacturer has ceased trading in the UK. Users should quarantine all remaining stock and dispose of all NGPod devices.
General SurgeryGi Hepatology Pancreatic Disorders1Full detail
Rectal catheters sold for personal use (gas and colic relievers): Do not use in infants, including for the treatment of colic, gas-related discomfort or constipation (DSI/2026/008)
GP
Do not use rectal catheters sold for personal use (gas and colic relievers) in infants, including for colic, gas-related discomfort or constipation. Stop using these products immediately and dispose of any unused devices.
General practiceGi Hepatology Pancreatic Disorders1Full detail
Class 2 Medicines Recall: Zentiva Pharma UK Limited, Fingolimod Zentiva 0.5 mg Capsules, EL(26)A/37
PharmacyGP
Zentiva UK Pharma Limited are recalling one batch of Fingolimod Zentiva 0.5 mg Capsules (BN 4L01372H) as a precautionary measure due to a potential risk of contamination with metal particles.
National Patient Safety Alert: ResMed Astral 100 and 150 Ventilators: Potential for Patient Harm due to Unexpected Interruption of Ventilation Therapy (NatPSA/2026/004/MHRA)
ResMed has issued a Field Safety Notice regarding Astral 100 and 150 ventilators and associated Printed Circuit Board Assembly spare parts manufactured prior to October 2024. A fault in the device can cause unexpected interruption of ventilation therapy.
Devices supplied without valid UKCA/CE conformity markings or certification: remove from use and place in quarantine (DSI/2026/007)
PharmacyGP
The MHRA has become aware of a number of medical devices being supplied on the UK market that do not comply with the requirements of the UK Medical Device Regulations 2002 (UK MDR 2002) or with the EU Medical Devices Regulation (EU) 2017/745 (EU MDR).
MHRA medicines recall classes describe urgency when published in the alert title. PharmAnalyser surfaces the class from the indexed title — it does not infer Class 1 for unclassified alerts.
Class 1
Highest urgency — immediate action typically required.
Class 2
Action required within 48 hours in most cases.
Class 3
Action required within one week in most cases.
Class 4
For information — action may be required at local discretion.